VA vs VV ECMO: Flow, Oxygenation and LV Loading

What is the difference between VA and VV ECMO?

VV ECMO returns oxygenated blood to the venous circulation, so the native heart still provides systemic pumping. VA ECMO returns blood to the arterial circulation and adds direct circulatory support. Neither pump flow nor one oxygen saturation value, alone, establishes effective oxygen delivery to every organ.

Key takeaways

  • VV ECMO supports gas exchange while native cardiac function remains essential.
  • VA arterial return does not guarantee left-ventricular unloading.
  • Regional oxygenation can differ during femoral VA support.

In this video

  1. Good oxygen numbers and poor perfusion
  2. Gas exchange and systemic pumping
  3. The extracorporeal circuit and sweep
  4. VV return and the native heart
  5. Indirect effects on right ventricular load
  6. Mixing hemoglobin and oxygen supply
  7. Recirculation
  8. Oxygen content and membrane transfer math
  9. VA return and circulatory support
  10. Femoral VA and left ventricular load
  11. Two aortic streams and regional oxygenation
  12. A VV patient with falling native output
  13. Femoral VA and two oxygen readings
  14. CRNA interview reasoning
  15. Quick checks

Why does VV ECMO still depend on the heart?

The membrane exchanges gas across separate blood and gas streams. In VV support, the treated blood returns to the venous side and then passes through the native heart. The circuit is not a direct substitute for systemic cardiac pumping.

Mixing and recirculation affect how much of the circuit’s work contributes to systemic oxygen delivery. A displayed pump flow is therefore different from a complete measurement of the patient’s effective oxygen delivery.

How does VA support change circulation?

Arterial return provides a direct contribution to systemic flow. With femoral VA return, retrograde aortic flow can also increase left-ventricular afterload. Supporting systemic pressure is not the same as guaranteeing that the LV is unloaded.

Native ejection, loading conditions and the circuit interact. Simple flow addition requires explicit assumptions and a clear definition of what each measurement includes. It should not become a universal rule for every patient or circuit configuration.

Why can oxygenation differ between regions?

During femoral VA support, native ejection from poorly oxygenating lungs can meet the circuit’s retrograde stream in the arterial circulation. The balance can change, so lower-body oxygenation does not establish what reaches the coronary or cerebral circulation.

Different regional oxygen contents are not, by themselves, proof that regional tissue hypoxia occurred. This lesson explains the mechanism and the limitation of single measurements; it does not prescribe sampling, sweep or pump adjustments.

Sources

Related guides

Frequently Asked Questions

Why does VV ECMO require native cardiac function?

It returns blood to the venous circulation. The native heart still has to move that blood through the systemic circulation.

Does VA ECMO necessarily unload the LV?

No. Peripheral arterial return can increase LV afterload. Direct circulatory support and unloading are different effects.

Can upper- and lower-body oxygenation differ on femoral VA ECMO?

Yes. Native ejection and retrograde circuit flow can create different regional oxygen contents. One lower-body measurement cannot represent every organ.